YS-8000C Defi-monitor/defibrillator
Physical characteristics:
Main unit: Portable
Dimensions:L320xW205xH410
Weight:9.8kg
Technical specifications
DEFIBRILLATOR.
(1) Defibrillator type Manual, Synchronized, Asynchronized,
(2) Out put waveform Monophasic technology
(3) Energy accuracy ∠+1%
(4) Energy selection 0, 3, 5, 7, 10, 20, 30, 50,100, 200, 300, 360 joule(normal at a resistance of 50)
(5) Charging time Maximum 8 seconds to 360 Joules
(6) Standard Adult/Pediatric Paddles Reusable extermal adult paddles(pediatric paddles integrated)
Power requirements.
(a) AC Power supply 110V, 60 Hz /220V,50 Hz
(b) Vehicle Voltage DC 12V
Internal Battery Backup
(1) Battery type Rechargeable lead-acid battery/ lithium battery(optional)
(2) Charging time Time-Minimum of 4 hours for full charging
(3) Battery charging During AC Power Supply operation of the device.
(4) Charging /Low Battery Indication Red LED
(5) Battery Capacity 100% of the battery will provide 120 mins of monitoring; and 30 discharges at 360 joules.
Display
-LED Dimensions 7’diagonal {155mm*88mm}
-Type High resolution true-color graphic LED
-Resolution 480X234
ECG
ECG lead: Ⅰ, Ⅱ, Ⅲ, AVR, AVL, AVF, V1~V6, GND
S-T segment analysis: -2.0~2.0mv
Arrhythmia analysis: yes
Protegtion: With stand 4000vAc/50Hz voltage in isolation and work against eletrosurgical interence and defibrillation
Respiration
Measurement method: Thoracic impedance
NIBP
Working mode: Manual/Automatic/stat(5 minutes)
Method: Automatic oscillation
Cufftype: standard Adult; optional Pediatric/Neonatal
Temperature
Measurement range 20~45℃
Resolution 0.1℃
Pulse Oximetry(spO2)
Display :spo2 value,pulse histongram,waveform,Pulse
Range: 0~99%for adult,pediatric and neonatal patients
Probe :standard adult finger clip
optional pediatric Y-type clip and neonatal wrap
Pulse range
Pulse range: 0-255bpm
Standard parameter & configuration
ECG/ RESP,/SPO2,/TEMP,/PR,/NIBP/DEFIBRILLATOR
Optional parameter
Recorder