Medical Anesthesia Machine
Model:YSAV6301D
Integrated Breathing Circuit
The integrated breathing circuit with an APL valve provides excellent tightness and safety.
The circuit components are resistant to high temperatures of 134°C and can withstand high-pressure disinfection.
Replacing the drive bellows is not required when operating for both adult and pediatric anesthesia.
High-precision built-in dual flow sensors are located at both the inhalation and exhalation ports.
The 2L sodium lime tank features a single-tank structure, making it more convenient to disassemble.
High Precision Vaporizer
Advanced automatic compensation for temperature, flow and pressure.
Used for low-flow anesthesia, saving more anesthetic drugs.
Enflurane, Isoflurane, Sevoflurane, Halothane etc, are available.
ACGO
International standard design. (diameter: 22mm)
Convenient for patients oxygen inhalation and postoperative recovery.
Compatible with T circuit or Bain circuit for open-operation.
AGSS (optional)
Made of aluminum alloy: high strength, lightweight and rust-resistant.
Provides a safety guarantee for the treatment of anesthetic waste gases in the operating room.
Technique Specifications
Physical Specifications
Dimensions: 70*85*140cm
Weight: 65kg
Casters: Front wheels with lock
Screen: 10.1 inch LCD / Touch screen
Working Specifications
Power: AC 100V-240V, 50Hz±2%
Pipeline supply
Gas configuration: O2, N2O, Air
Pipeline input range: 280-600kPa
Battery Power
Run-time: ≥180mins
Pre-setting Functions
Language: English
Calibration: Automatic
Applications
Patient range: Adult/Pediatric
Ventilator Specifications
Modes of Ventilation
VCV: Volume Control Ventilation
AC: Assist / Control Ventilation
SIMV: Synchronized Intermittent Mandatory Ventilation
MANUAL: Manual
PCV: Pressure Control Ventilation
PSV: Pressure Support Ventilation
STAND-BY: STAND-BY
Ventilator Parameter
Working type: Electronically controlled, pneumatic driven
Tidal volume range: 20-1500ml
Pressure range: 5-60cmH2O
Respiration rate range: 1-100bpm(1-40bpm under SIMV)
I:E range: 4:1-1:8
Ptrigger range: -20-20cmH2O(based on PEEP)
Flow trigger: 0-15L
Minute volume range: >18L/Min
SIGH range: 0-5/100
System Standard
Hypoxic guard system: N2O cut-off valve, O2 concentration >25%
Safety valve: <12.5kPa
Fresh gas compensation: 25-75L/min
Volume of CO2 absorber: 2L
Flow meter: 5 tubes, Cascade 0.1-15L/min 02, 0.1-15L/min N2O, 0.1-15L/min Air
Vaporizer: 2 Selectatec, maximum 2 mounts
(lsoflurane/Enflurane/Sevoflurane/Halothane)
Monitoring
Display graphics: Waveforms of P-T, F-T, V-T, loops of P-V, V-F, F-P
Types: VTi, VTe, MV, Pmax, Pplateau, Ri, Cs, Freq, FiO2, ETCO2(Optional)
Alarm Limits
Audio/Visual: No tidal volume, MV, Paw, FiO2 limitation O2, Air failure, AC power failure, Battery low
PEEP
Type: Integrated, Electronic control
Range: OFF, 3-30cmH2O
Optional features
1.AGSS
2.Suction Unit