We use cookies to improve your online experience. By continuing browsing this website, we assume you agree our use of cookies.
Categories

Trolley-mounted Anesthesia Workstation YSAV8834

Model: YSAV8834
  • ModelYSAV8834
  • BrandYSENMED
  • PriceGet Price
  • MOQ1
  • Ship FromFrom Guangzhou,China
  • ShippingAir/Sea
  • -
    +
    Categories:Anesthesia Machines
    Share:
    Description Get Price Review
    Trolley-mounted Anesthesia Workstation
    Model:YSAV8834
    YSAV8834 anaesthesia machine is electronically-controlled pneumatic-driven. More comfortable, More safe, More humanization design and easy to operate. The Breathing circuit is PPSU special material that can withstand 134°C high temperature and high pressure steam sterilization. Also, YSAV8834 anaesthesia machine can extra ETCO2 and AGSS.
    Patient Type
    Adult, child
    Ventilation Mode
    VCV, SIMV, PCV, MAN, Standby
    Main Setting Parameters
    Tidal volume: 0, 20~1500ml
    Respiratory rate: 0,1~100 bpm
    Breathing ratio: 1:10~10:1
    Inspiratory time: 0.5~5.0S
    Pressure trigger: 0~20cmH2O
    Flow trigger: 0~20L/min
    PEEP: OFF, 4~30 cmH2O
    Pressure control Pmax: 5~70 cmH2O
    SIMV respiratory rate: 0, 4~20 bpm
    Main Monitoring Parameters
    Inspiratory tidal volume: 0~2000 mL
    Expiratory tidal volume: 0~2000 mL
    Minute ventilation: 0~30 L/min
    Respiratory rate: 0~100 bpm
    FiO2: 18vol%~100vol%
    ETCO2: 0~13vol%
    AC power failure: 
    When the AC power is cut off,it will be automatically converted to battery power
    supply, and an audible and visual alarm will be issued.
    Basic Parameters
    Air Source: O2, N2O, AIR
    Loops: Pressure-Volume, Flow rate-Volume
    Waveforms: Pressure-Time, Flow rate-Time
    Evaporators: Enflurane, lsoflurane, Sevoflurane, Halothane
    Apnea alarm time: 10~40S
    Low Battery alarm: ≥15min
    Run out of battery alarm: ≥3min
    Alarm mute timing: ≤120S